The MIC600OZ measures ozone gas concentration using dual-path UV absorption with an LED light source. Its 19-inch 3U chassis provides a rack-mounted configuration.
Features
- Automatic zero calibration device
- Temperature and pressure compensation
- Ozone concentration, temperature and pressure display
- Built-in flowmeter, ozone exhaust catalyst and inlet filter
Technical Specifications
| Parameter | Specification |
|---|---|
| Measurement principle | Dual-path UV absorption |
| Standard range | 0–300 g/Nm³ |
| Alternative ranges | 0–200, 0–100 or 0–50 g/Nm³; 0–1000 ppm |
| Display | 5-inch touchscreen |
| Display resolution | 0.01 g/Nm³ |
| Sample gas flow | 0.5 ± 0.2 L/min |
| Inlet pressure | <0.1 MPa |
| Sampling | Positive pressure; optional negative pressure |
| Communication | RS-485 |
| Analog output | Active 4–20 mA |
| Relays | High and low alarm point outputs |
| Power supply | AC 110–220 V |
| Ambient temperature | −20 to 50 °C |
| Dimensions (H × W × D) | 133 × 482 × 300 mm; 19-inch 3U chassis |
Alternative ranges are configuration options. Display resolution is not measurement accuracy.
Application and Selection
Use for generator outlet or exhaust gas concentration measurement with suitable sample handling. Exhaust gas measurement requires a dehumidification device.